CR10 and VW Sensors, 2002/12/18 1
Introduction
About This Manual
This manual tells how to read VW sensors using Campbell Sci-
entific CR10 and CR10X data loggers with the AVW1 or
AVW100 interface and the AM416 or AM16/32 multiplexer.
CR10 and CR10X
Instructions
The CR10 provides instruction 28 to read VW sensors. The
CR10X provides instruction 28 and instruction 131 to read VW
sensors. Both loggers use the instruction 4 to read thermistors
or RTDs.
VW Interface Modules There are three vibrating wire interfaces modules available, the
AVW1, the AVW4, and the AVW100. We discuss only the
AVW1 and AVW100. The AVW1 can be used with either
instruction 28 or instruction 131. The AVW100 works only with
instruction 131.
Multiplexer The older AM416 multiplexer has been replaced by the AM16/
32 multiplexer. Either multiplexer can be used to control 16 VW
sensors with temperature or 32 VW sensors without tempera-
ture.
Logger Software Logger software called PC208 or LoggerNet is used to create and
transfer a monitoring program to the logger and later to retrieve
stored data from the logger. You can find sample monitoring
programs at www.slopeindicator.com. Go to Support - Tech
Notes - Sample Programs for CR10X.
You can modify these programs to meet your needs. If you need
help, you can request programming services from Slope Indica-
tor. Note that Slope Indicator cannot provide unpaid support for
these programs.
About
Data Reduction
VW sensors return a frequency signal that must be converted to
units of pressure, strain, or displacement. The conversion pro-
cess and other data reduction issues are discussed in the manual
for each type of sensor.