Campbell Scientific CS511-L Sensorex Dissolved Oxygen Probe
Operating Conditions:
Temperature
Pressure:
Minimum Submersion Depth:
Minimum Water Flow:
0° to 50°C (32° to 122°F)
0 to 100 psig
60 mm (2 ½ in)
5 cm/s (2 in/sec) across membrane
Calibration: In air or in air saturated water
Temperature Compensation: Automatic from 4° - 40°C (40° - 104°F)
Range of Dissolved Oxygen: 0.5 to 50 ppm
Probe Electrolyte: NaCl + glycerol (prevents freezing)
3. Shipping Kit and Accessories
Except for the agitator, these items are for Sensorex’s
DO6400/T. Refer to Appendix B if you have a DO6200/T.
NOTE
3.1 Shipping Kit
(1) Membrane replacement tool
(1) Bottled DO electrolyte, 250 ml
(2) Teflon membranes
(2) Membrane o-rings
(2) Membrane spaces
3.2 Optional Probe Accessories
PT4 Agitator for stagnant conditions
22261 Maintenance Kit containing (5) Teflon membranes, (5) membrane o-
rings, (5) tensioning washers, and a 250-ml bottle of electrolyte
22262 Maintenance Kit containing (25) Teflon membranes, (25) membrane
o-rings, (25) tensioning washers, and a 500-ml bottle of electrolyte
22263 Spare Parts Kit containing (2) membrane locks, (2) tensioning washers,
(2) body o-rings, and (1) membrane replacement tool
4. Optional Agitator
The PT4 Agitator is a reliable and robust agitator for use in conjunction with
probes subjected to bio-fouling in ponds and stagnant water conditions.
O
2
probes require a minimum water velocity across their membranes to function
properly. Therefore, to measure DO in stagnant water conditions, it is necessary to
move the water past the membrane to get accurate and reliable DO measurements.
In many instances the water also has a high bio-loading and the probes become
fouled resulting in inaccurate DO measurements.