Campbell Scientific CS511-L Sensorex Dissolved Oxygen Probe
3
3. Optional Probe Accessories
Agitator for stagnant conditions.
Membrane replacement tool.
DO electrolyte, 500 ml.
Teflon membranes, 5 pack.
Teflon membranes, 25 pack.
4. Optional Agitator
The PT4 Agitator is a reliable and robust agitator for use in conjunction with
probes subjected to bio-fouling in ponds and stagnant water conditions.
O
2
probes require a minimum water velocity across their membranes to function
properly. Therefore, to measure DO in stagnant water conditions, it is necessary
to move the water past the membrane to get accurate and reliable DO
measurements. In many instances the water also has a high bio-loading and the
probes become fouled resulting in inaccurate DO measurements.
FIGURE 4-1. Preventing Bio-fouling of the CS510-L
The PT4 Agitator overcomes these problems. The device is designed so that a
soft bristle brush sweeps across the probe membrane or sensor tip. This sweeping
action of the brush provides the required water velocity as well as prevents the
membrane from becoming bio-fouled.
The optimum sweeping frequency depends upon the design of probe and type of
membrane used and water conditions. An ON-time of 0.25 seconds and OFF-
time of 5 seconds is suitable in most circumstances.
Agitator’s overall size: 7-1/8" x 3-1/4" diameter, 1-1/4 lb. (180mm x 83mm, 0.6 kg)
Supplied with 10 ft. (3 meters) cable.
Power required: 10.5 to 18 VDC at the agitator, 1.1 amps. Maximum ON-time is
3 seconds.