Campbell HMP 35A User Manual

Browse online or download User Manual for Unknown Campbell HMP 35A. Campbell Scientific HMP35c

  • Download
  • Add to my manuals
  • Print
  • Page
    / 14
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 0
MODEL HMP35C TEMPERATURE AND
RELATIVE HUMIDITY PROBE
INSTRUCTION MANUAL
REVISION: 1/99
COPYRIGHT (c) 1990-1999 CAMPBELL SCIENTIFIC, INC.
Page view 0
1 2 3 4 5 6 ... 13 14

Summary of Contents

Page 1 - INSTRUCTION MANUAL

MODEL HMP35C TEMPERATURE ANDRELATIVE HUMIDITY PROBEINSTRUCTION MANUALREVISION: 1/99COPYRIGHT (c) 1990-1999 CAMPBELL SCIENTIFIC, INC.

Page 2

HMP35C TEMPERATURE AND RH PROBE8TABLE 8. Wiring for Example 3Description Color CR10(X)Temperature Orange SE 3 (2H)Relative Humidity Green SE 4 (2L)Te

Page 3 - MODEL HMP35C

HMP35C TEMPERATURE AND RH PROBE9;Turn the HMP35C on.;05: Do (P86)1: 41 Set Port 1 High;Yellow wire (C1);Pause 150 milliseconds, before making the meas

Page 4 - 3. INSTALLATION

HMP35C TEMPERATURE AND RH PROBE10TABLE 9. Wiring for Example 4Description Color CR10(X)Temperature Orange SE 3 (2H)Relative Humidity Green SE 4 (2L)T

Page 5

HMP35C TEMPERATURE AND RH PROBE11;Compute the saturation vapor pressure in kPa. The temperature must be in degrees Celsius.;06: Saturation Vapor Pres

Page 6 - 5. EXAMPLE PROGRAMS

HMP35C TEMPERATURE AND RH PROBE12TABLE 11. Temperature, Resistance, andDatalogger OutputTemperature(°C)Resistance(Ω)Output(°C)-40.00 4067212 -39.18-3

Page 7 - 6. LONG LEAD LENGTHS

WARRANTY AND ASSISTANCEThe HMP35C TEMPERATURE AND RELATIVE HUMIDITY PROBE is warranted by CAMPBELLSCIENTIFIC, INC to be free from defects in materials

Page 8

1MODEL HMP35CTEMPERATURE AND RELATIVE HUMIDITY PROBE1. GENERAL DESCRIPTIONThe HMP35C Temperature and RelativeHumidity probe contains a thermistor for

Page 9 - ENVIRONMENTS

HMP35C TEMPERATURE AND RH PROBE2FIGURE 1. Error Produced by Polynomial Fit to Published Values3. INSTALLATIONThe HMP35C must be housed inside a radi

Page 10

HMP35C TEMPERATURE AND RH PROBE3FIGURE 3. HMP35C with UT018 Mounting Bracket and Crossarm andUT12VA Radiation Shield Mounted on a UT30 TowerFIGURE 4.

Page 11 - 8. ABSOLUTE HUMIDITY

HMP35C TEMPERATURE AND RH PROBE44. WIRINGConnections to Campbell Scientific dataloggersare given in Table 3. The probe is measured bytwo single-ende

Page 12

HMP35C TEMPERATURE AND RH PROBE5;Pause 150 milliseconds, before making the measurement, so the;probe can stabilize on the true relative humidity.;03:

Page 13 - 10. MAINTENANCE

HMP35C TEMPERATURE AND RH PROBE6Example 2. Sample CR10(X) Program measuring HMP35C Temperatureusing DC Half Bridge with Delay;Measure the HMP35C ther

Page 14 - 11. REFERENCES

HMP35C TEMPERATURE AND RH PROBE7;Measure the HMP35C relative humidity.;07: Volt (SE) (P1)1: 1 Reps2: 5 2500 mV Slow Range;CR500 (2500 mV); CR23X (1000

Comments to this Manuals

No comments