TDR100
29
water content = 9.5%
water content = 25%
FIGURE 8-2. Waveforms collected in a sandy loam using CS610 probe
with RG8 connecting cable. Volumetric water content values are 10,
16, 18, 21 and 25%. Solution electrical conductivity is 1.0 dS m
-1
.
water content = 18%
water content = 37%
FIGURE 8-3. Waveforms collected in a sandy loam using CS610 probe
with RG8 connecting cable. Volumetric water content values are 10,
18, 26, 30 and 37%. Solution electrical conductivity is 10.2 dS m
-1
.
The combined effect of long cable runs and high soil electrical conductivity
must be considered when TDR measurements are taken.