24
2: 3 >=
3: 2 F Change as required
4: 31 Exit Loop if True
45: End (P95)
46: Set Port(s) (P20)
1: 0 C8..C5 = 0/0/0/0
2: 0 C4..C1 = 0/0/0/0
47: Beginning of Loop (P87)
1: 1 Delay
2: 0 Loop Count
48: Timer (P26)
1: 99 Loc [ _________ ]
49: IF (X<=>F) (P89) Wait 15 s before sampling for cross overs
1: 99 X Loc [ _________ ]
2: 3 >=
3: 15 F Change as required
4: 15 Set Flag 5 High
50: If Flag/Port (P91)
1: 15 Do if Flag 5 is High
2: 31 Exit Loop if True
51: End (P95)
52: Do (P86)
1: 22 Set Flag 2 Low
53: End (P95)
54: Batt Voltage (P10) Store battery voltage in loc (98)
1: 98 Loc [ _________ ]
End Program
5.8. Operation of the logger program - C21-HP4.DLD
Here we describe the HPV program for a CR21X logger and the normal tests that you
may want to do. We also describe some 'normal' program behaviour that may be expected
during the course of a measurement period. Note: After you have finished playing with
the logger, ALWAYS press *0 to recompile any changes, or to continue by leaving the
logger in the 'LOG1' mode.
5.8.1. To change the interval between heat pulses (alter Inst-1)
enter *1 1 A A A x A *0 where x = interval in minutes