The default time is 30 minutes
5.8.2. To fire a test heat-pulse
enter *6 A D 1 (the lights should go on, and the program will run)
enter *6 A D 6 for an early exit of program
enter *0 to continue
5.8.3. To check if the system is working properly (generally a good idea!)
fire a test heat pulse (see 5.8.2)
check the initial voltages (+/- 20 mV is OK, -9999 is stuffed)
i.e. enter *6 x A where x = 41..56
check the present voltages (+/- 100 mV is OK, -9999 is stuffed)
i.e. enter *6 x A where x = 21..36
enter *0 to continue
Note. Readings are in mV so anything in the range +/- 200 is OK.
5.8.4. To examine
t
z
data (stored in locations 1..16)
enter *6 x A where x is the location.
loc. 1 = tz1-1 (outer) .... loc. 4 = tz1-4 (inner)
loc. 5 = tz2-1 (outer) .... loc. 8 = tz2-4 (inner)
loc. 9 = tz3-1 (outer) .... loc.12 = tz3-4 (inner)
loc. 13 = tz4-1 (outer) .... loc.16 = tz4-4 (inner)
enter *0 to continue
5.8.5. To find out the total running-time since the heat-pulse was fired (loc. 99)
enter *6 99 A
enter *0 to continue
5.8.6. To change the duration of the heat pulse (alter Inst-48)
enter *1 48 A A A x A *0 where x is duration in s
default value is 2.0 s (this is also the maximum time)
5.8.7. To change the delay between firing a heat pulse and testing for a 'cross-over'. This
may be necessary for really fast flows (alter Inst-56)
enter *1 56 A A A x A *0 where x is delay in multiples of 0.1s
default value is 15 s, i.e. x=150
5.8.8. To change the maximum time for a
t
z
calculation (alter Inst-15)