CS526 ISFET pH Probe2/12Copyright © 2009-2012Campbell Scientific, Inc.
CS526 ISFET pH Probe Maximum Cable Length: 100 m (328 ft) Cable Type: Three-twisted pair, 24 awg cable with Santoprene® jacket Sensor Material: PEEK
CS526 ISFET pH Probe instruction (see Section 8.2.2.2). Table 8-1 shows the “get data” command and its response. TABLE 8-1. Get Data Command and Res
CS526 ISFET pH Probe 8.2.2 Instruction Descriptions 8.2.2.1 SerialOpen() Instruction The SerialOpen() instruction has the following syntax: SerialOp
CS526 ISFET pH Probe 8.2.2.4 SplitStr Instruction The SplitStr() instruction has the following syntax: SplitStr(SplitResult,SearchString,FilterStrin
CS526 ISFET pH Probe TABLE 9-1. Calibration Standards pH CSI Part Number 4 25587 7 25586 10 25588 Frequency of calibration depends on the level of
CS526 ISFET pH Probe 11. Maintenance Campbell Scientific recommends that the CS526 be cleaned and calibrated periodically to insure accurate readings
CS526 ISFET pH Probe Diaphragm Chip FIGURE 11-1. Tip of the CS526 Probe 10
Appendix A. Calibration This calibration process uses 7 pH and 4 pH buffer solutions. Protect the sensing chip from UV radiation during calibration.
Appendix A. Calibration A-2 Following is a simple example program to facilitate the two point calibration. 'CR1000 Series Datalogger 'Defin
Campbell Scientific Companies Campbell Scientific, Inc. (CSI) 815 West 1800 North Logan, Utah 84321 UNITED STATES www.campbellsci.com • info@campbel
Warranty The CS526 ISFET pH Probe is warranted for six (6) months subject to this limited warranty: “PRODUCTS MANUFACTURED BY CAMPBELL SCIENTIFIC, INC
Assistance Products may not be returned without prior authorization. The following contact information is for US and international customers residing
CS526 Table of Contents PDF viewers: These page numbers refer to the printed version of this document. Use the PDF reader bookmarks tab for links to
CS526 Table of Contents ii Appendix A. Calibration... A-1 Figure 11-1. Tip of the CS526
CS526 ISFET pH Probe 1. Introduction The CS526 ISFET pH Probe measures pH from 0 to 14 in aqueous or semi-solid solutions. It outputs TTL serial dat
CS526 ISFET pH Probe 4. Quickstart For complete installation, programming, and calibration information, see Sections 7 through 9. The CS526 is shippe
CS526 ISFET pH Probe 5. Overview The submersible CS526 is designed to reliably provide accurate temperature compensated pH measurements. The pH-sens