CSAT3 THREE DIMENSIONAL SONIC ANEMOMETER
10
hygrometer, the thermocouple and hygrometer
data must be shifted back in time two triggers.
This can be done on-line with a datalogger.
10. DATALOGGER PROGRAMMING
USING SDM
10.1 SDM-CSAT3 (P107) USED WITH CR23X,
21X, OR CR10(X)
SDM-CSAT3 (P107) is an Input/Output
instruction that controls and retrieves data from
the CSAT3 (see Table 10-1). This instruction is
available on the CR23X, 21X (with CSAT3
PROMs, P/N 10006), the CR10X, and the CR10
(with library special PROMs).
TABLE 10-1. SDM-CSAT3 (P107) Instruction
Parameters
PARAMETER DATA DEFAULT
NUMBER TYPE DESCRIPTION
01: 2 Reps
02: 2 SDM Address
03: 2 Option
04: 4 Ux Input Location
Reps: The number of CSAT3s on the SDM bus
that will be sequentially addressed with the
SDM-CSAT3 (P107) instruction.
SDM Address: The unique SDM Address of the
CSAT3 that the datalogger will communicate
with. The CSAT3 address is entered as a base
four number in parameter 2 (see Table 3-1).
The SDM Address is sequentially incremented if
the Reps option is greater than 1.
The wind data and the speed of sound (sonic
virtual temperature) is scaled by Instruction 107
into m s
-1
or degrees Celsius.
CAUTION: SDM address 33 (base 4) is a
special address used during a Group
Trigger. Do not program the datalogger to
address SDM address 33.
Option 1, 2, 3, 5, 6, 10, 12, 15, 20, 30, and 60:
“Set Execution Parameter” tells the CSAT3
which measurement parameters to use and
what frequency to expect the measurement
trigger from the datalogger. The datalogger
Execution Interval and the appropriate CSAT3
Execution Parameter is summarized in Table
10-2 and 10-3.
Option 61 and 62: "Set Execution Parameter"
tells the CSAT3 to oversample (Section 8.1.2)
the wind measurements at a rate of 60 Hz and
expect a trigger at a rate of 10 Hz (option 61) or
20 Hz (option 62) from a CR23X or 21X
datalogger.
Option 51 and 52: "Set Execution Parameter"
tells the CSAT3 to oversample (Section 8.1.2)
the wind measurements at a rate of 48 Hz and
expect a trigger at a rate of 8 Hz (option 51) or
16 Hz (option 52) from a CR10(X) datalogger.
Option 0, 90: “Trigger and Get Wind & c Data”
sends a measurement trigger to the CSAT3
whose SDM address is specified in parameter 2.
The CSAT3 also sends data to the datalogger.
Option 91: “Trigger and Get Wind & T
s
Data”
sends a measurement trigger to the CSAT3
whose SDM address is specified in parameter
2. The CSAT3 also sends wind speed and the
sonic virtual temperature (defined in Equation
(9) in Appendix C) via the SDM bus to the
datalogger.
Option 92: “Trigger and Get Wind & c-340
Data” sends a measurement trigger to the
CSAT3 whose SDM address is specified in
parameter 2. The CSAT3 also sends wind
speed and the speed of sound minus 340 m s
-1
via the SDM bus to the datalogger.
Option 97: “Get Wind and c-340 Data After a
Group Trigger”. The CSAT3, specified in
parameter 3, sends wind speed and the speed
of sound minus 340 m s
-1
to the datalogger. A
new measurement is not triggered. The SDM-
CSAT3 (P107) Instruction must be preceded by
the Group Trigger (P110) Instruction to use this
option.
Option 98: “Get Wind and Ts Data After a
Group Trigger”. The CSAT3, specified in
parameter 3, sends the wind speed and sonic
virtual temperature (defined in Equation (9) in
Appendix C) to the datalogger. A new
measurement is not triggered. The SDM-
CSAT3 (P107) Instruction must be preceded by
the Group Trigger (P110) Instruction to use this
option.
Option 99: “Get Wind and c Data After a
Group Trigger”. The CSAT3, specified in
parameter 3, sends the wind speed and speed
of sound to the datalogger. A new
measurement is not triggered. The SDM-
CSAT3 (P107) Instruction must be preceded by
the Group Trigger (P110) Instruction.