CSAT3 THREE DIMENSIONAL SONIC ANEMOMETER
11
TABLE 10-2. CR23X and 21X Execution Interval and CSAT3 Execution Parameter
CR23X/21X Execution
Interval (seconds)
CR23X/21X Execution
Frequency (Hz)
CSAT3 Execution Parameter
0.05 20 20
0.0625 16 15
0.075 13.333 12
0.0875 to 0.1 11.429 to 10 10
0.1125 to 0.1625 8.889 to 6.154 6
0.175 to 0.2 5.714 to 5 5
0.2125 to 0.325 4.706 to 3.077 3
0.3375 to 0.5 2.963 to 2 2
0.5125 to 1 1.951 to 1 1
TABLE 10-3. CR10(X) Execution Interval and CSAT3 Execution Parameter
CR10(X) Execution Interval
(seconds)
CR10(X) Execution Frequency (Hz) CSAT3 Execution Parameter
0.0625 16 15
0.078125 12.8 12
0.09375 10.667 10
0.109375 to 0.15625 9.143 to 6.4 6
0.171875 to 0.1875 5.818 to 5.333 5
0.203125 to 0.328125 4.923 to 3.047 3
0.34375 to 0.5 2.909 to 2 2
0.515625 to 1 1.939 to 1 1
Ux Input Location: The Input Location to store
the first of five data values. The CSAT3 returns
the following data after receiving the “Trigger
and Get Data” or “Get Data after a Group
Trigger” command in parameter 3. These data
are: u
x
, u
y
, u
z
, c or T
s
, and a diagnostic word,
where u
x
, u
y
, and u
z
are the wind speeds along
the x, y, and z axes defined by the anemometer
head (see Figure 4-1), c is the speed of sound,
and T
s
is the sonic virtual temperature. The
wind speed and the speed of sound are in units
of m s
-1
, and T
s
is in degrees Celsius. If the
Reps option in parameter 1 is greater than 1,
the number of sequential Input Locations used
is Reps 5.
10.2 DIAGNOSTIC WORD
The fifth output from the CSAT3 is the diagnostic
word. The diagnostic word describes the status
of the anemometer and is useful in filtering data.
For example, when the anemometer is locking in
on the ultrasonic signals after receiving the
Acquire signals command, the -99999’s it sends
to the datalogger, for the science data, can be
filtered out of the statistics based on the
diagnostic word, see the example flux program
in Section A.
The SDM-CSAT3 instruction (P107), decodes
diagnostic word and represents it in base 10.
The bits in the diagnostic word are grouped
together into three categories. Bits 15 through
12 are four diagnostic flags (see Table 10-4 and
10-5), bits 11 through 6 represent the range the
measurements are reported in (used by
Instruction 107), and bits 5 through 0 are a
counter. A detailed description of decoding the
diagnostic word is given in Appendix B.
TABLE 10-4. Diagnostic Word
Diagnostic Word
b15 - b12 b11 b10 b09 b08 b07 b06 b05 - b00
diagnostic
flags
u
x
range
u
y
range
u
z
range
counter
The following are four specific Diagnostic
Words (Base 10) and the situation that
produces them.