APPENDIX D. SDM COMMUNICATIONS AND LONG SIGNAL CABLES
D-2
parameters that were tested and their failure
rate as a percent of attempts.
The following report was created from the
above SPLIT parameter file, SDM.PAR.
TABLE D-2. SDM Clock Rate Report
SDM Clock Rate Report
SDM Failures Attempts
Param % Attmpt
1 100 1
4 100 1
7 100 1
9 100 17
10 96 450
11 0 449
12 0 449
13 0 432
From the above results, the fastest rate that the
SDM bus could run, without encountering
communication errors, would be using a value
of 11 for the SDM clock parameter in Instruction
Set SDM Clock (P115). However, to have a
safety margin, a value of 12 is used for the SDM
parameter. The test showed that using a value
of 12 or the SDM parameter in Instruction Set
SDM Clock did not cause the CR23X to overrun
the Tables each scan, thus the SDM clock rate
is left at parameter 12.
With the CR9000, collect the datatable “report”.
This datatable contains an SDM Clock Rate
Report similar in form to the one generated by
SPLIT (Table D-2).
Instruction Set SDM Clock (P115) with a value
of 12 for the SDM parameter is added to the
flux program and tested.