CSAT3 THREE DIMENSIONAL SONIC ANEMOMETERINSTRUCTION MANUALREVISION: 3/98COPYRIGHT (c) 1996-1998 CAMPBELL SCIENTIFIC, INC.

Page 2 - WARRANTY AND ASSISTANCE

CSAT3 THREE DIMENSIONAL SONIC ANEMOMETER54.3 LEVELINGOver flat level terrain, adjust the anemometerhead so that the bubble within the level is in the

Page 3 - TABLE OF CONTENTS

CSAT3 THREE DIMENSIONAL SONIC ANEMOMETER6TABLE 5-6. CSAT3 Analog Output WiringDescription Colorux sig high Brownux sig low Blackuy sig high Reduy sig

Page 4 - APPENDICES

CSAT3 THREE DIMENSIONAL SONIC ANEMOMETER7TABLE 6-2. Calibration for Analog OutputsSlopeLow Range(m s-1 V-1)SlopeHigh Range(m s-1 V-1)Offset(m s-1)ux6

Page 5

CSAT3 THREE DIMENSIONAL SONIC ANEMOMETER8internal trigger then it will take measurements atthis rate until it is changed.Note that the power consumpti

Page 6 - 2. SPECIFICATIONS

CSAT3 THREE DIMENSIONAL SONIC ANEMOMETER9FIGURE 8-1. Typical Average CSAT3 Power Requirements at +12 VDCMiMi + 1Mi + 2Mi + 3Pi - 1Si - 2PiSi - 1Pi +

Page 7 - 3. INITIAL SETUP

CSAT3 THREE DIMENSIONAL SONIC ANEMOMETER10hygrometer, the thermocouple and hygrometerdata must be shifted back in time two triggers.This can be done o

Page 8 - 4. INSTALLATION

CSAT3 THREE DIMENSIONAL SONIC ANEMOMETER11TABLE 10-2. CR23X and 21X Execution Interval and CSAT3 Execution ParameterCR23X/21X ExecutionInterval (seco

Page 9 - 10.00 cm

CSAT3 THREE DIMENSIONAL SONIC ANEMOMETER12-99999, Anemometer does not Respond.This is the special case when the dataloggerattempts to address an anemo

Page 10 - 5. WIRING

CSAT3 THREE DIMENSIONAL SONIC ANEMOMETER13EXAMPLE 1. Measuring the CSAT3 Usingthe SDM Output and a CR23X Datalogger;{CR23X};;Flag 1 - Set low to conf

Page 11 - 6. CSAT3 OUTPUTS

CSAT3 THREE DIMENSIONAL SONIC ANEMOMETER14There are no user serviceable components onthe CSAT3. The only time the electronics boxmust be opened is to

Page 12 - 8. CSAT3 TRIGGER

WARRANTY AND ASSISTANCEThe CSAT3 THREE DIMENSIONAL SONIC ANEMOMETER is warranted by CAMPBELL SCIENTIFIC,INC. to be free from defects in materials and

Page 13

A-1APPENDIX A. EXAMPLE FLUX PROGRAMA copy of the example CR23X program is available on the Campbell Scientific ftp site atftp://ftp.campbellsci.com/p

Page 14 - 9. DATA PIPELINE DELAY

APPENDIX A. EXAMPLE FLUX PROGRAMA-2TABLE A-1. CR23X Sensor ConnectionsChannel Sensor Color1H HMP45C (temperature) Yellow1L HMP45C (temperature) Purp

Page 15 - USING SDM

B-1APPENDIX B. SERIAL COMMUNICATIONS WITH THE CSAT3B.1 CSAT3 RS-232 COMMANDSThe CSAT3 powers up its RS-232 drivers when it detects that the RTS (req

Page 16

APPENDIX B. SERIAL COMMUNICATIONS WITH THE CSAT3B-21 The CSAT3 will ignore this command if a datalogger (SDM) is triggering the anemometer.2 After th

Page 17 - 61440, Lost Trigger

APPENDIX B. SERIAL COMMUNICATIONS WITH THE CSAT3B-3set RTS line lowrepeat until finished collecting dataprocess datasend W or S commandread wind data

Page 18

APPENDIX B. SERIAL COMMUNICATIONS WITH THE CSAT3B-4TABLE B-5. Converting uz Wind from Word 2Range Word 4 Range Code Conversion Output Resolutionb07

Page 19 - 12. CALIBRATION

APPENDIX B. SERIAL COMMUNICATIONS WITH THE CSAT3B-5B.5 CSAT3 RS-232 STATUS FORMATThe CSAT3 returns a 10 ASCII byte status in response to an S or P c

Page 20

C-1APPENDIX C. CSAT3 MEASUREMENT THEORYC.1 THEORY OF OPERATIONC.1.1 WIND SPEEDEach axis of the CSAT3 pulses two ultrasonicsignals in opposite direc

Page 21

APPENDIX C. CSAT3 MEASUREMENT THEORYC-2where Ts is sonic virtual temperature and γd isthe ratio of specific heat of dry air at constantpressure to th

Page 22 - B.1 CSAT3 RS-232 COMMANDS

D-1APPENDIX D. SDM COMMUNICATIONS ANDLONG SIGNAL CABLESD.1 BRIEF DESCRIPTION OF SDMCLOCK RATESThe fastest rate that a datalogger and SDMsensors can

Page 23

ICSAT3 THREE DIMENSIONAL SONIC ANEMOMETERINSTRUCTION MANUALTABLE OF CONTENTSPAGE1. GENERAL...

Page 24

APPENDIX D. SDM COMMUNICATIONS AND LONG SIGNAL CABLESD-2parameters that were tested and their failurerate as a percent of attempts.The following repo

Page 25 - B.4 WORD 4

TABLE OF CONTENTSIIAPPENDICESA. EXAMPLE FLUX PROGRAM...A-1B.

Page 26

TABLE OF CONTENTSIIIB-5 Converting uz Wind from Word 2...B-4B

Page 27

1CSAT3 THREE DIMENSIONAL SONIC ANEMOMETER1. GENERALThe CSAT3 is a three dimensional sonicanemometer. It measures wind speed and thespeed of sound on

Page 28

CSAT3 THREE DIMENSIONAL SONIC ANEMOMETER2DIGITAL RS-232:Baud rate: 9600Data type: 2-byte integerSDM AND RS-232 DIGITALRANGE AND RESOLUTION:Full scal

Page 29 - LONG SIGNAL CABLES

CSAT3 THREE DIMENSIONAL SONIC ANEMOMETER3TABLE 3-1. SDM Addressing SchemeDataloggerAddress(base 4)CSAT3 ThumbSwitch(HEX)Address(base 10)00 0 001 1 10

Page 30

CSAT3 THREE DIMENSIONAL SONIC ANEMOMETER410.00 cmYZXLogan, Utah CSAT33-D SONIC ANEMOMETERS/N: 102MADE IN USA60oFIGURE 4-1. CSAT3 Coordinate System an